Install the hollow cathode lamp to be measured and adjust the wavelength of spectrophotometer.

Your current location: home >> News >> Company news

Install the hollow cathode lamp to be measured and adjust the wavelength of spectrophotometer.

date:2017-11-30 author: click:

test

1. install the hollow cathode lamp to be measured, adjust the lamp current and wavelength to the desired value.

2. ignition, setting instrument test parameters;

3. The manufacturer of atomic absorption spectrophotometer inserts the capillary into the deionized water, adjusts zero, inserts the sample capillary into the solution, after the absorbance shows stable, records the test results, inserts the capillary into the deionized water, returns to zero, and then determines in turn.

Turn off the machine

1. after the test, clean the atomizer for several minutes in the ignition state with clean deionized water.

2. close the main valve of the gas cylinder and turn off the gas valve of the instrument after burning the remaining gas in the pipeline.

3. release the gas and fuel gas knob on the instrument panel and turn the lamp current to zero.

4. turn off the instrument power supply and turn off the regulated power supply.

5. turn off fans and cooling water.

6. loose the gas cylinder pressure relief valve.

7. order off gas, air compressor, cathode lamp. Reset knob, turn off light multiplier tube negative high voltage power supply, total power supply.

8. wipe the burn seam with filter paper.

9. cleaning instrument, reset, fill in the instrument usage record.

Operation Rules of Atomic Absorption Spectrophotometer: Turn on the computer - turn on the main power supply - double-click the icon for self-examination - wait for self-examination after the completion of the element lamp selection - according to the wizard prompt for peak search operation.

Operating procedures for atomic absorption spectrophotometer:

Turn on the computer - turn on the mainframe power - double-click on the icon for self-test - wait for the completion of self-test element lamp selection - according to the wizard prompt for peak search operation.

Flame method: Click on the icon to set the sample - click on the icon to set the parameters - click on the icon to adjust the position of the atomizer - turn on the air compressor power supply (pressure 0.25-0.3 MPa) - turn on the acetylene cylinder switch (pressure adjustment to about 0.06MPa) - click on the icon - click on the icon to balance energy automatically. Click click icon to start the measurement. Turn off acetylene, air compressor and turn off main engine after measurement.

Graphite furnace method: click on the icon to set the sample - click on the icon to set the parameters - click on the icon to adjust the position of the atomizer - Open argon (pressure 0.6-0.8 MPa) - cooling water - graphite furnace power - click on the furnace body to open, install the graphite tube, click on the furnace body after installation - click on the icon to adjust energy - click on the settings Set up temperature program. Click on empty burn three times, click to start measuring. After measuring, turn off argon, cooling water, graphite furnace power and turn off the main engine.